Authors: Jakub Rafl, Filip Kulhanek, Petr Kudrna, Vaclav Ort, Karel Roubik
Citation
Rafl, J., Kulhanek, F., Kudrna, P., Ort, V. and Roubik, K., 2018. Response time of indirectly accessed gas exchange depends on measurement method. Biomedical Engineering/Biomedizinische Technik, 63(6), pp.647-655.
Fulltext in PDF & fulltext download
Download fulltext in PDF here: Response time of indirectly accessed gas exchange depends on measurement method
Published in Biomedical Engineering / Biomedizinische Technik
Abstract
Noninvasive techniques are routinely used for assessment of tissue effects of lung ventilation. However, comprehensive studies of the response time of the methods are scarce. The aim of this study was to compare the response time of noninvasive methods for monitoring of gas exchange to sudden changes in the composition of the inspired gas. A prospective experimental study with 16 healthy volunteers was conducted. A ventilation circuit was designed that enabled a fast change in the composition of the inspiratory gas mixture while allowing spontaneous breathing. The volunteers inhaled a hypoxic mixture, then a hypercapnic mixture, a hyperoxic mixture and finally a 0.3% CO mixture. The parameters with the fastest response to the sudden change of O2 in inhaled gas were peripheral capillary oxygen saturation (SpO2) and regional tissue oxygenation (rSO2). Transcutaneous oxygen partial pressure (tcpO2) had almost the same time of reaction, but its time of relaxation was 2–3 times longer. End-tidal carbon dioxide (EtCO2) response time to change of CO2 concentration in inhaled gas was less than half in comparison with transcutaneous carbon dioxide partial pressure (tcpCO2). All the examined parameters and devices reacted adequately to changes in gas concentration in the inspiratory gas mixture.
References
[1] Barker SJ, Curry J, Redford D, Morgan S. Measurement of carboxyhemoglobin
and methemoglobin by pulse oximetry: a human volunteer study. Anesthesiology 2006; 105: 892–897.
[2] Brugger H, Sumann G, Meister R, et al. Hypoxia and hypercapnia during respiration into an artificial air pocket in snow: implications for avalanche survival. Resuscitation 2003; 58: 81–88.
[3] Buehler S, Jensen MC, Lozano-Zahonero S, et al. The dynamics of carbon dioxide equilibration after alterations in the respiratory rate. Physiol Meas 2013; 34: 1151–1161.
[4] Cairo JM, Pilbeam SP. Mechanical ventilation: physiological and clinical applications. 5th ed. St. Louis: Elsevier 2012.
[5] Casati A, Squicciarini G, Malagutti G, Baciarello M, Putzu M, Fanelli A. Transcutaneous monitoring of partial pressure of carbon dioxide in the elderly patient: a prospective, clinical comparison with end-tidal monitoring. J Clin Anesth 2006; 18: 436–440.
[6] Cherniack NS, Longobardo GS, Staw I, Heymann M. Dynamics of carbon dioxide stores changes following an alteration in ventilation. J Appl Physiol 1966; 21: 785–793.
[7] Ehrenfeld JM, Cannesson M, editors. Monitoring technologies in acute care environments: a comprehensive guide to patient monitoring technology. New York: Springer 2014.
[8] Eichhorn L, Erdfelder F, Kessler F, et al. Evaluation of near-infrared spectroscopy under apnea-dependent hypoxia in humans. J Clin Monit Comput 2015; 29: 749–757.
[9] Feiner JR, Rollins MD, Sall JW, Eilers H, Au P, Bickler PE. Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia. Anesthesia 2013; 117: 847–858.
[10] Gehring H, Hornberger C, Matz H, Konecny E, Schmucker P. The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia. Resp Care 2002; 47: 48–60.
[11] Genbrugge C, Dens J, Meex I, et al. Regional cerebral oximetry during cardiopulmonary resuscitation: useful or useless? J Emerg Med 2016; 50: 198–207.
[12] Goldstein M. Carbon monoxide poisoning. J Emerg Nurs 2008; 34: 538–542.
[13] Górska K, Korczyński P, Maskey-Warzęchowska M, Chazan R, Krenke R. Variability of transcutaneous oxygen and carbon dioxide pressure measurements associated with sensor location.
Adv Exp Med Biol 2015; 858: 39–46.
[14] Hampson NB. Noninvasive pulse CO-oximetry expedites evaluation and management of patients with carbon monoxide poisoning. Am J Emerg Med 2012; 30: 2021–2024.
[15] Huttmann SE, Windisch W, Storre JH. Techniques for the measurement and monitoring of carbon dioxide in the blood. Ann Am Thorac Soc 2014; 11: 645–652.
[16] Kaymak C, Basar H. Carbon monoxide intoxication – Review. Fabad J Pharm Sci 2010; 35: 163–172.
[17] Kesten S, Chapman KR, Rebuck AS. Response characteristics of a dual transcutaneous oxygen/carbon dioxide monitoring system. Chest 1991; 99: 1211–1215.
[18] Lermuzeaux M, Meric H, Sauneuf B, et al. Superiority of transcutaneous CO2 over end-tidal CO2 measurement for monitoring respiratory failure in nonintubated patients: a pilot study. J Crit Care 2016; 31: 150–156.
[19] MacLeod DB, Cortinez LI, Keifer JC, et al. The desaturation response time of finger pulse oximeters during mild hypothermia. Anaesthesia 2005; 60: 65–71.
[20] Nishiyama T, Nakamura S, Yamashita K. Effects of the electrode temperature of a new monitor, TCM4, on the measurement of transcutaneous oxygen and carbon dioxide tension. J Anesth 2006; 20: 331–334.
[21] Nishiyama T, Nakamura S, Yamashita K. Comparison of the transcutaneous oxygen and carbon dioxide tension in different electrode locations during general anaesthesia. Eur J Anaesth 2006; 23: 1049–1054.
[22] Prockop LD, Chichkova RI. Carbon monoxide intoxication: an updated review. J Neurol Sci 2007; 262: 122–130.
[23] Redford D, Paidy S, Kashif F. Absolute and trend accuracy of a new regional oximeter in healthy volunteers during controlled hypoxia. Anesthesia 2014; 119: 1315–1319.
[24] Reid CW, Martineau RJ, Miller DR, Hull KA, Baines J, Sullivan PJ. A comparison of transcutaneous, end-tidal and arterial measurements of carbon dioxide during general anaesthesia. Can J Anaesth 1992; 39: 31–36.
[25] Restrepo RD, Hirst KR, Wittnebel L, Wettstein R. AARC clinical practice guideline: transcutaneous monitoring of carbon dioxide and oxygen: 2012. Resp Care 2012; 57: 1955–1962.
[26] Rich K. Transcutaneous oxygen measurements: implications for nursing. J Vasc Nurs 2001; 19: 55–59.
[27] Roth D, Herkner H, Schreiber W, et al. Accuracy of noninvasive multiwave pulse oximetry compared with carboxyhemoglobin from blood gas analysis in unselected emergency department patients. Ann Emerg Med 2011; 58: 74–79.
[28] Roubik K, Sieger L, Sykora K. Work of breathing into snow in the presence versus absence of an artificial air pocket affects hypoxia and hypercapnia of a victim covered with avalanche snow: a randomized double blind crossover study. PLoS One 2015; 10: e0144332.
[29] Scheeren TWL, Schober P, Schwarte LA. Monitoring tissue oxygenation be near infrared spectroscopy (NIRS): background and current applications. J Clin Monit Comput 2012; 26: 279–287.
[30] Sivarajan VB, Bohn D. Monitoring of standard hemodynamic parameters: heart rate, systemic blood pressure, atrial pressure, pulse oximetry, and end-tidal CO2. Pediatr Crit Care Med 2011; 12: S2–11.
[31] Spelten O, Fiedler F, Schier R, Wetsch WA, Hinkelbein J. Transcutaneous PtcCO2 measurement in combination with arterial blood gas analysis provides superior accuracy and reliability in ICU patients. J Clin Monit Comput 2017; 31: 153–158.
[32] Storre JH, Magnet FS, Dreher M, Windisch W. Transcutaneous monitoring as a replacement for arterial PCO2 monitoring during nocturnal non-invasive ventilation. Resp Med 2011; 105: 143–150.
[33] Tobias JD. Transcutaneous carbon dioxide monitoring in infants and children. Pediatr Anesth 2009; 19: 434–444.
[34] Trivedi NS, Ghouri AF, Shah NK, Lai E, Barker SJ. Pulse oximeter performance during desaturation and resaturation: a comparison f seven models. J Clin Anesth 1997; 9: 184–188.
[35] van Oppen JD, Daniel PS, Sovani MP. What is the potential role of transcutaneous carbon dioxide in guiding acute noninvasive ventilation? Resp Care 2015; 60: 484–491.
[36] Vazquez AL, Fukuda M, Tasker ML, Masamoto K, Kim SG. Changes in cerebral arterial, tissue and venous oxygenation with evoked neural stimulation: implications for hemoglobinbased functional neuroimaging. J Cereb Blood Flow Metab 2010; 30: 428–439.
[37] Walsh BK, Crotwell DN, Restrepo RD. Capnography/capnometry during mechanical ventilation: 2011. Resp Care 2011; 56: 503–509.
[38] Yip WL. Evaluation of the clinimetrics of transcutaneous oxygen measurement and its application in wound care. Int Wound J 2015; 12: 625–629.
[39] Yu M, Morita SY, Daniel SR, Chapital A, Waxman K, Severino R. Transcutaneous pressure of oxygen: a noninvasive and early detector of peripheral shock and outcome. Shock 2006; 26: 450–456.